Setting up, commissioning, operation
3.4.6 Operating hour display
Fig. 30: Operating hour display
At switch-on (normal status):
The LEDs light up continuously in
green:
The LEDs flash continuously in
green:
The LEDs flash continuously in red:
50
The operating hour display is located below the On-Off
switch and consists of two LEDs.
The LEDs indicate the following states:
The LEDs flash for a few seconds.
The LEDs then switch to continuous green light.
The unit is in normal status and working as intended.
The last 100 hours of the lamp operating hours have
been reached (purchase replacement lamp).
The maximum number of lamp operating hours has been
reached (measured on the basis of the operating hours
counted).
The lamp must be replaced.
Operating time at switch-on
Every time the unit is switched on, two hours
of operating time are added to the operating
hours, as every switching operation damages
the lamps.
The unit was designed for 24-hour operation
and can be operated continuously.
Avoid unnecessary switching operations.