Fig. 31: Position the Torx screws
Fig. 32: Removing the plug-in
base
Fig. 33: Removing the lamp from
the holder
Installing lamps
3.
Release and remove the fixing screws from the side
wall.
Two of the screws are Torx screws (TX25) (Fig. 31)
and need to be removed with the Torx screw-
driver provided.
4.
Remove the side wall.
5.
Only handle the UV lamps by the ceramic ends.
6.
Pull the plug-in base out of the UV lamp.
Hold the plug-in base by the socket, and not by the
cable!
7.
Grip the UV lamp by the ceramic ends and pull it out of
the holder.
ð The UV lamps have now been fully removed.
NOTICE!
The units may only be operated with the UV
lamps listed in
for units and replacement parts' on page 58.
SteriWhite Air Q330/Q600
Servicing, maintenance and cleaning
Ä Chapter 3.8 'Ordering data
English
53