English
Servicing, maintenance and cleaning
Fig. 34: UV lamp (top), plug-in
base with retaining pin (bottom)
Fig. 35: Position the Torx screws
54
Wear clean, non-linting gloves.
1.
Connect the plug-in base.
Make sure that the retaining pin is inserted into the
correct recess (Fig. 34).
2.
Place the new UV lamps in the holder.
CAUTION!
Pinching hazard
You can pinch your fingers when inserting the
side panel.
– Slowly and carefully place the side panel
on the device.
– Do not reach under the side wall, but hold
it by the edge.
3.
Place the side wall on the unit.
4.
Insert the fixing screws and tighten them by hand.
Two of the screws are Torx screws (TX25), and must
be installed in the positions shown in Fig. 35.
To tighten these two screws, you will need the Torx
screwdriver provided.
5.
(Optional) You can now reinstall the unit on the wall or
ceiling.
6.
Then reset the operating hours (see
'Reset operating hours' on page 56).
SteriWhite Air Q330/Q600
Ä Chapter 3.5.4