WARNING: THIS DEVICE SUPPLIES
HIGH-CONCENTRATION OXYGEN THAT
PROMOTES RAPID BURNING. DO NOT
ALLOW SMOKING OR OPEN FLAMES
WITHIN THE SAME ROOM OF (1) THIS
DEVICE, OR (2) ANY OXYGEN-CAR-
RYING ACCESSORY. FAILURE TO
OBSERVE THIS WARNING CAN RESULT
IN SEVERE FIRE, PROPERTY DAMAGE
AND / OR CAUSE PHYSICAL INJURY OR
DEATH.
WARNING: DO NOT USE YOUR OXYGEN CON-
CENTRATOR IN THE PRESENCE OF FLAMMABLE
GASES. THIS CAN RESULT IN RAPID BURNING
CAUSING PROPERTY DAMAGE, BODILY INJURIES
OR DEATH.
WARNING: DO NOT LEAVE A NASAL CANNULA
ON CLOTHING, BED COVERINGS OR CHAIR
CUSHIONS. IF THE UNIT IS TURNED ON BUT NOT
IN USE, THE OXYGEN WILL MAKE THE MATERI-
AL FLAMMABLE. SET THE I/0 POWER SWITCH
TO THE 0 (OFF) POSITION WHEN THE OXYGEN
CONCENTRATOR IS NOT IN USE.
WARNING: USE NO OIL, GREASE, OR PETRO-
LEUM-BASED OR OTHER FLAMMABLE PROD-
UCTS WITH THE OXYGEN-CARRYING ACCESSO-
RIES OR THE OXYGEN CONCENTRATOR. ONLY
WATER BASED, OXYGEN COMPATIBLE LOTIONS
OR SALVES SHOULD BE USED. OXYGEN AC-
CELERATES THE COMBUSTION OF FLAMMABLE
SUBSTANCES.
WARNING: THE USE OF SOME OXYGEN ADMIN-
ISTRATION ACCESSORIES NOT SPECIFIED FOR
USE WITH THIS OXYGEN CONCENTRATOR MAY
IMPAIR ITS PERFORMANCE. RECOMMENDED
ACCESSORIES ARE REFERENCED WITHIN THIS
MANUAL
CAUTION: Always operate the unit in an
upright position.
NewLife Intensity 10
Proper Setting of Oxygen Flowmeter
To set the proper flow of supplemental oxygen, turn
the flowmeter adjustment knob left or right until the
ball inside the flowmeter centers on the flow line
number prescribed by your physician.
To view the flowmeter at the proper angle, note that
the back line and the front numbered line must give
the appearance of just one line.
CAUTION: Normally, you should
not need to adjust the flowmeter on
your unit. If you turn the flowmeter
adjustment knob clockwise, you will
decrease and can shut off the flow
of oxygen from your unit. For your
convenience, the flowmeter is marked in
½ LPM increments. For units equipped
with the 2 LPM flowmeter option,
the flowmeter is marked in 1/8 LPM
increments for flow settings up to 2
LPM.
CAUTION: The Oxygen Concentrator may be
used during sleep under the recommendation of a
licensed clinician.
PN MN239-C4 F | User Manual ENG - 11