NewLife Intensity 10
Limited oxygen flow.
Condensation collects in the
oxygen tubing when you use
the humidifier bottle.
Intermittent alarm sounds.
NewLife Intensity 10
displays alarm and produces
intermittent beep.
Oxygen concentrator does
not turn on.
All other problems.
20 - ENG PN MN239-C4 F | User Manual
Dirty or obstructed humidifier
bottle.
Defective nasal cannula,
face mask, catheter, and/or
oxygen delivery tube, or other
accessory.
Other leak or restriction.
Unit not properly ventilated.
Elevated operating
temperature.
Equipment malfunction.
Refer to Alarm Conditions
table.
Not connected to external
power.
General malfunction.
Remove the humidifier bottle (if used) from
the oxygen outlet. If flow is restored, clean or
replace with a new humidifier bottle.
Remove nasal cannula, face mask, or other
accessories from oxygen tubing. If proper flow is
restored, replace with new nasal cannula, face
mask, or other accessories.
Disconnect delivery tubing at oxygen outlet
(front of unit). If proper flow is restored, check
oxygen tubing for kinks or obstructions. Replace
if needed.
Contact your Equipment Provider.
Make sure unit is positioned away from curtains
or drapes, hot air registers, heaters, and
fireplaces. Be certain to place the unit so all
sides are at least 12 inches (30.5 cm) away from
a wall or other obstruction. Do not place the unit
in a confined area.
Allow oxygen tubing to dry out, or replace with
new tubing. Refill humidifier bottle with COLD
water. DO NOT OVERFILL.
Set I/0 power switch to 0 position, use your
reserve oxygen supply and consult your
Equipment Provider immediately.
Refer to Alarm Conditions table.
Power the unit through the outlet.
Ensure that external connects are secure.
Contact your Equipment Provider, and change to
another source of oxygen as necessary.
Set I/0 power switch to the 0 position, use
your reserve oxygen supply and consult your
Equipment Provider immediately.