NewLife Intensity 10
WARNING: USE ONLY WATER-BASED LOTIONS
OR SALVES THAT ARE OXYGEN COMPATIBLE
PRIOR TO OR DURING OXYGEN THERAPY. NEVER
USE PETROLEUM OR OIL-BASED LOTIONS OF
SALVES TO AVOID THE RISK OF FIRE OR BURNS.
WARNING: DO NOT USE LIQUID DIRECTLY ON
THE UNIT. A LIST OF UNDESIRABLE CHEMICAL
AGENTS INCLUDES BUT IS NOT LIMITED TO THE
FOLLOWING: ALCOHOL AND ALCOHOL-BASED
PRODUCTS, CONCENTRATED CHLORINE-BASED
PRODUCTS (ETHYLENE CHLORIDE), AND
OIL-BASED PRODUCTS (PINE-SOL
THESE ARE NOT TO BE USED TO CLEAN THE
PLASTIC HOUSING ON THE OXYGEN CONCEN-
TRATOR, AS THEY CAN DAMAGE THE UNIT'S
PLASTIC. CLEAN THE CABINET, CONTROL PANEL,
AND POWER CORD ONLY WITH A MILD HOUSE-
HOLD CLEANER APPLIED WITH A DAMP CLOTH
(NOT WET) OR SPONGE, AND THEN WIPE ALL
SURFACES DRY. DO NOT ALLOW ANY LIQUID TO
GET INSIDE THE DEVICE.
Note: Always follow the cannula manufacturer's
instructions for proper use. Replace the disposable
cannula as recommended by the cannula manufactur-
er or your equipment provider. Additional supplies are
available from your equipment provider.
Note: The Manufacturer does not recommend the
sterilization of this equipment.
Filters
At least one time each week, wash the air intake
gross particle filter, which is located in the back
of the unit. Your Equipment Provider may advise
you to clean it more often, depending upon your
operating conditions. Follow these steps to properly
clean the air intake filter:
Note: Do not operate the unit without the intake gross
particle filter in place.
1. Remove the filter and wash it in a warm solution
of soap and water.
2. Rinse the filter thoroughly, and remove excess
water with a soft, adsorbent towel. Ensure that the
filter is completely dry before replacing it.
3. Replace the dry filter.
18 - ENG PN MN239-C4 F | User Manual
Reserve Oxygen Supply
Your Equipment Provider may recommend another
source of supplemental oxygen therapy in case
there is a mechanical failure or a power outage.
, LESTOIL
).
®
®
Cannula Replacement
Always follow the cannula manufacturer's
instructions for proper use. Replace the nasal
cannula or oxygen tubing as recommended by the
cannula manufacturer or your oxygen provider.
Your physician or oxygen provider will provide you
with cleaning and replacement instructions.
Additional supplies for replacement are available
through your oxygen provider.
CAUTION: The Manufacturer recom-
mends an alternate source of supple-
mental oxygen in the event of a power
outage, alarm condition, or mechanical
failure. Consult your physician or Equip-
ment Provider for the type of reserve
system required.