5. SELECT PATIENT INTERFACE
The AIRVO 2 can be used with a variety of patient interfaces. Read the separate user instructions for the
patient interface that will be used, including all warnings.
Nasal cannula
Optiflow™+
Optiflow™
OPT942
OPT842
OPT944
OPT844
OPT946
OPT846
All patient interfaces are Type BF applied parts.
The following table shows the target dew-point temperature settings and target flow settings able to be used
with these interfaces.
Patient Interface
Low temperature ambient conditions may prevent the unit from reaching a 37 °C target temperature
setting at high target flow settings. In these cases, consider decreasing the target flow setting.
At altitude, the maximum flow rates achievable may be lower than those in the above table, by
approximately 5 L/min per 1000 m (3000 ft).
WARNINGS
To avoid burns:
• Do not modify the breathing tube or interface in any way.
• Do not use any patient interfaces not listed here.
4. INSTALL HEATED BREATHING TUBE
One end of the heated breathing tube has a blue plastic sleeve. Lift the sleeve
and slide the connector onto the unit. Push the sleeve down to lock.
WARNINGS
To avoid burns:
• Do not modify the breathing tube or interface in any way.
• Do not allow the breathing tube to remain in direct contact with skin for prolonged
periods of time. The healthcare professional shall assess the conditions for safe
contact, such as duration and skin condition.
• Do not add heat above ambient levels to any part of the breathing tube or interface
e.g. by covering with a blanket or by heating with infrared radiation, an overhead
heater, or an incubator.
• Do not use an insulating sleeve or any similar accessories which are not recommended
by Fisher & Paykel Healthcare.
CAUTIONS
• Position the heated breathing tube away from any electrical monitoring leads (EEG,
ECG/EKG, EMG, etc), to minimize any possible interference with the monitored signal.
Optiflow™ Junior/Junior 2
OPT316/OPT318/
OJR416/OJR418
(Refer to "Using AIRVO 2" -
"Junior Mode")
2
OPT942
OPT316/OJR416
2
OPT944
OPT318/OJR418
OPT942
OPT946
OPT944
OPT870
RT013
OPT946
OPT970
OPT980
OPT842
OPT844
OPT846
OPT870
RT013
A – 5
Tracheostomy
Mask interface adapter
interface
OPT970 / OPT870
OPT980 / RT013 (with mask)
Note that the OPT980/RT013 Mask
Interface Adapter is designed to be
used with vented masks only. Do not
use sealed masks.
10
10
10
10
10
10
10
10
10
10