Fitting And Application - Ottobock 5065N Instrucciones De Uso

Ocultar thumbs Ver también para 5065N:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 19
Only clean the product with the approved cleaning agents.
NOTICE
Contact with oils, salves, lotions or other products that contain oils or acids
Insufficient stabilisation due to loss of material functionality
Do not expose the product to oils, salves, lotions or other products that contain oils or acids.
5 Handling
INFORMATION
The daily duration of use and period of application are generally determined by the physi­
cian.
The initial fitting and application of the product must be carried out by qualified personnel.
Instruct the patient in the handling and care of the product.
Instruct the patient to see a physician immediately if any exceptional changes are noted (e.g.
worsening of the complaint).
5.1 Size Selection
1) Measure the chest circumference.
2) Determine the orthosis size (see size chart).

5.2 Fitting and Application

CAUTION
Incorrect or excessively tight application
Skin irritation due to overheating, local pressure points due to tight application to the body
Ensure that the product is applied properly and fits correctly.
When in doubt, do not continue to use the product if skin irritation occurs.
NOTICE
Use of a worn or damaged product
Limited effectiveness
Before each use, check the product for functional reliability and for possible wear or dam­
age.
Do not continue using a product that is no longer functional, or that is worn or damaged.
1) Open all hook-and-loop closures.
2) Guide the arm through the shoulder cuff (see fig. 2).
3) Apply the shoulder cuff to the shoulder (see fig. 3).
4) Fasten the hook-and-loop closure on the upper arm strap (see fig. 4). In doing so, make sure
the red seam is below the shoulder joint gap.
5) Bring the chest strap around the chest (see fig. 5). In doing so, position the axilla pad under
the axilla.
6) Fasten the chest strap buckle on the shoulder cuff.
7) Adjust the length of the chest strap with the hook-and-loop closure (see fig. 6). To do this,
shorten the chest strap with scissors and fasten it with an Y-hook-and-loop.
8) Guide the forearm through the forearm cuff (see fig. 7). In doing so, the snap fasteners on the
forearm cuff have to be on top.
9) Make sure that the snap fasteners of the forearm strap are close to the pivot point of the elbow
joint without allowing the orthosis to cover the olecranon (see fig. 8). In doing so, the closure
on the forearm cuff has to be at the front.
10

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

5065n omo neurexa plus