Table 2
Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the
user of the device should assure that it is used in such an environment .
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC61000-4-5
Voltage dips,
short interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
Power frequency
(50Hz/60Hz)
magnetic field
IEC 61000-4-8
NOTE
U
is the a.c. mains voltage prior to application of the test le vel.
T
30
IEC 60601 test level
±8 kV contact
±15 kV air
power supply lines:
±2 kV
input/output lines:
±1 kV
line(s) to line(s): ±1 kV
line(s) to earth: ±2 kV
100 kHz repetition
frequency
0%U
0.5 cycle
T;
At 0°, 45°, 90°, 135°,
180°,225°,270° and 315°
0%U
; 1 cycle
T
and
70%U
; 25/30 cycles
T
Single phase: at 0°
0% U
; 300 cycle
T
30 A/m
50Hz/60Hz
Compliance level
±8 kV contact
±15 kV air
power supply lines:
±2 kV
line(s) to line(s): ±1 kV
100 kHz repetition
frequency
0% U
; 0.5 cycle
T
At 0°, 45°, 90°, 135°,
180°,225°,270° and
315°
0% U
; 1 cycle
T
and
70% U
; 25/30 cycles
T
Single phase: at 0°
0% U
;300 cycle
T
30 A/m
50Hz/60Hz
Electromagnetic
environment - guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.