REUSE PRECAUTION STATEMENT
For single patient use only. Do not reuse, reprocess or resterilize. Reuse, reprocessing or resterilization
may compromise the structural integrity of the device and/or lead to device failure which, in turn, may
result in patient injury, illness or death. Reuse, reprocessing or resterilization may also create a risk of
contamination of the device and/or cause patient infection or cross-infection, including, but not limited
to, the transmission of infectious disease(s) from one patient to another. Contamination of the device may
lead to injury, illness or death of the patient.
POTENTIAL COMPLICATIONS
Possible complications that may result from use of this device include, but are not limited to: hematoma,
recurrent bleeding, local venous thrombosis, nerve damage, pain or numbness, complex regional pain
syndrome, allergic reaction, and artery occlusion.
INSTRUCTIONS FOR USE
Device Placement
1. Ensure site is clean and dry.
2. For left hand placements use catalog code SDRB-REG-LT. For right hand placements use catalog code
SDRB-REG-RT.
3. For a large patient wrist/hand, you may add the extension band (included) prior to placing the
PreludeSYNC DISTAL (FIGURE 1).
4. Aspirate the sheath, then withdraw the sheath approximately one inch (2-3 cm).
5. Place the band on the hand/wrist with band #1 on the little nger side, band #2 on the thumb side, and
band #3 facing up toward the thumb (FIGURE 2).