6. Place the center of the balloon "target" over the arteriotomy (location where the sheath entered the
artery, approximately 1-2mm proximal to the skin puncture site) (FIGURE 3).
7. Fasten band #1 and #2 tightly around the wrist, then wrap band #3 around the saddle of the thumb
without any slack (FIGURE 4).
NOTE: If the sheath is in the way of securing band section #3, then band section #3 may be secured
after removal of sheath.
8. Fill the syringe (included) with 10mL of air.
NOTE: Maximum balloon ll volume: 10mL
9. Engage the syringe into the valve on the top of the device.
10. Slowly inflate the balloon with air while simultaneously removing the sheath. Once the sheath
is completely removed, continue to inject air into the balloon until bleeding has stopped.
NOTE: While in ating, maintain pressure on syringe plunger to avoid any inadvertent release of air.
11. Adjust air volume in balloon to achieve hemostasis. If bleeding or hematoma are present, additional
air may be injected until bleeding stops.