Smith & Nephew Renasys go Manual De Usuario página 14

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 86
Clearing therapy time
Therapy time must be cleared between each patient
use by entering clinician mode as follows:
1.
Press and hold POWER button for 2 seconds to turn
device off.
2.
Simultaneously press and hold DOWN, SELECT and
POWER buttons for 2 seconds. Screen will display
"Continuous" and "Intermittent" on two lines.
3.
Use DOWN button to scroll to "Clear Time",
press SELECT button.
4.
The screen will display "Clear Active Time?",
press SELECT button.
5.
Use DOWN button to scroll to "Yes", press SELECT
button.
6.
The screen will display "Active Time Cleared" for
5 seconds and return to main menu displaying
"Continuous" and "Intermittent" on two lines.
7.
Press and hold POWER button for 2 seconds to turn
device off.
Total time of operation
RENASYS™ GO devices display total number of
accumulated hour's device has been used for therapy.
Total time can be viewed by entering clinician mode as
follows:
1.
Press and hold POWER button for 2 seconds to turn
off device.
2.
Simultaneously press and hold DOWN, SELECT and
POWER buttons for 2 seconds. Screen will display
"Continuous" and "Intermittent" on two lines.
3.
Use DOWN button to scroll to "Total Time", press
SELECT button.
4.
Screen will display "Total Time" followed by a value
indicating the number of accumulated hours device
has delivered therapy.
5.
After 5 seconds, screen will clear and return to main
menu displaying "Continuous" and "Intermittent" on
two lines.
6.
Press and hold POWER button for 2 seconds to turn
device off.
14

Publicidad

Tabla de contenido
loading

Tabla de contenido