Intermittent mode
1.
Press and hold POWER button for 2 seconds to turn
off device.
2.
Simultaneously press and hold DOWN, SELECT and
POWER buttons for 2 seconds. Screen will display
"Continuous" and "Intermittent" on two lines.
3.
Use DOWN button to scroll to "Intermittent".
4.
Press SELECT button to confirm selection. Screen
will display "Welcome, Starting", current software
version (e.g., V 1.06), and change to display
"Standby" and therapy set-point.
5.
Set stopwatch to zero.
6.
Adjust therapy set-point to 80mmHg using UP and
DOWN buttons. Press SELECT button to start device
and simultaneously start stopwatch.
7.
Verify device operates, status/alarm indicator
illuminates solid green, screen displays
"Intermittent" and therapy set-point.
8.
Verify device operates for 5 minutes ± 30 seconds
and then stops operating.
9.
Once device stops, reset stopwatch to zero and
start stopwatch. Verify device does not operate for
2 minutes ± 30 seconds, status/alarm indicator
remains illuminated solid green, and screen
displays "Intermittent" and "--- mmHg".
10.
Press SELECT to stop device. Verify device stops
operating, status/alarm indicator is no longer
illuminated, and screen displays "Standby" and
therapy set-point.
11.
Pause stopwatch and return device to "Continuous"
mode.
Keypad lock check
1.
Press SELECT button to start device with set-point
at 80mmHg. Verify device starts operating, status/
alarm indicator illuminates solid green, and screen
displays "Continuous" and therapy set-point.
2.
Press and hold KEYPAD LOCK button for 2 seconds.
Verify indicator light above button illuminates
solid blue, screen displays "Keypad Locked", and
changes to "Continuous" and therapy set-point.
3.
Press SELECT button to attempt to stop device.
Verify device does not stop operating and screen
displays "Keypad Locked".
4.
Press and hold KEYPAD LOCK button for 2 seconds.
Verify indicator light above button is no longer
illuminated and the screen displays "Keypad Active".
5.
Press SELECT button to stop device.
9