Smith & Nephew Renasys go Manual De Usuario página 27

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 86
Guidance and manufacturer's declaration – electromagnetic emissions RENASYS™ GO (REF 66801496)
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should
assure that it is used in such an environment.
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-2
WARNING: The device should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use
is necessary, the device should be observed to verify normal operation in the configuration in which it will be used.
Recommended separation distances between portable and mobile RF communications equipment and the device
The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer
or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the device as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum output
power of transmitter
(W)
0.01
0.1
1.0
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Compliance
Group 1
Class B
Class A
Comply
Separation distance according to frequency of transmitter (m):
150 kHz to 80 MHz
d = 1.2√P
0.12
0.38
1.2
3.8
12
Electromagnetic environment – guidelines
The device uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause interference in nearby electronic equipment.
The device is suitable for use in all establishments other than
domestic and those directly connected to the public low-
voltage power supply network that supplies buildings used
for domestic purposes.
80 MHz to 800 MHz
d = 1.2√P
0.12
0.38
1.2
3.8
12
27
800 MHz to 2.5 GHz
d = 2.3√P
0.23
0.73
2.3
7.3
23

Publicidad