GE Uniflux 10 Instrucciones De Funcionamiento página 203

Tabla de contenido

Publicidad

Step
4
5
Nested block: Start and stop the
product transfer
Step
1
2
3
Instruction: Transfer feed pump
Step
1
2
UniFlux 10 Instrucciones de funcionamiento 28993587 AE
Action
In the Text pane, select the End_Block instruction of the Block DELAY_3_SEC
block.
In the Instruction box:
Set Breakpoint to 0.05 min (equals 3 seconds).
Click the Change button.
Action
In the Text pane, select the Base Time instruction of the Block
PRODUCT_TRANSFER block.
In the Method Editor menu, select Block:New.
In the New Block dialog:
In the Name field, type an appropriate name, for instance
PRODUCT_TRANSFER_VOLUME.
In the Base group box, select Time.
In the Call group box, select From: Block PRODUCT_TRANSFER and At:
0.00 min.
Click OK to insert the new block.
The new block will be placed inside the Block PRODUCT_TRANSFER block.
Action
In the Text pane, select the Base Time instruction of the Block
PRODUCT_TRANSFER_VOLUME block.
In the Instruction box:
Set Breakpoint to 0.00 min.
In the Instructions group box, select Recirc.
Select TransferFeedPump, set Flow to 90%, click Insert.
A Create a method with UNICORN 5
A.2 An example of using the Method Editor
A.2.4 Block 2: Product transfer
203

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido