Instruction: Set the tank level
watch
Step
1
2
3
4
Set block end time
Change the block end time so that the tank level has time to reach the volume set in the
tank level watch.
Step
1
2
3
UniFlux 10 Instrucciones de funcionamiento 28993587 AE
Action
In the Text pane, select the Base Time instruction of the Block
TMP_START_STOP block.
In the Instruction box:
Set Breakpoint to 0.00 min.
•
In the Instructions group box, select Watch and Watch_FIT142VTot.
•
In the Parameters field:
Set Test to Greater_Than.
•
Set Value to an appropriate permeate volume.
•
Set Action to END_BLOCK.
•
Click Insert.
Action
In the Text pane, select the End_Block instruction of the Block
TMP_START_STOP block.
In the Instruction box, set Breakpoint to a suitable time, for instance 60.00
min.
Click Change.
A Create a method with UNICORN 5
A.2 An example of using the Method Editor