A Create a method with UNICORN 5
A.2 An example of using the Method Editor
A.2.5 Block 3: Fed-batch concentration
Step
3
Instruction: Setting Delta P value
Step
1
2
3
4
Instruction: Setting Delta P
variations
Step
1
2
212
Action
In the New Block dialog:
In the Name field, type an appropriate name, for instance
•
STABLE_DELTA_P.
In the Base group box, select Time.
•
In the Call group box, select From: Block FED_BATCH_CONCENTRATION
•
and At: 0.00 min.
Click OK to insert the new block.
•
The new block will be placed inside the Block FED_BATCH_CONCENTRATION
block.
Action
In the Text pane, select the Base Time instruction of the Block
STABLE_DELTA_P block.
In the Instruction box:
Set Breakpoint to 0.00 min.
•
In the Instructions group box, select Recirc and DeltaP.
•
In the Parameters field, set Setpoint to an appropriate value, for instance
1.5 bar.
Click Insert.
Action
In the Text pane, select the Base Time instruction of the Block
STABLE_DELTA_P block.
In the Instruction box:
Set Breakpoint to 0.00 min.
•
In the Instructions group box, select Specials and DeltaP_WatchPar.
•
UniFlux 10 Instrucciones de funcionamiento 28993587 AE