GE Uniflux 10 Instrucciones De Funcionamiento página 212

Tabla de contenido

Publicidad

A Create a method with UNICORN 5
A.2 An example of using the Method Editor
A.2.5 Block 3: Fed-batch concentration
Step
3
Instruction: Setting Delta P value
Step
1
2
3
4
Instruction: Setting Delta P
variations
Step
1
2
212
Action
In the New Block dialog:
In the Name field, type an appropriate name, for instance
STABLE_DELTA_P.
In the Base group box, select Time.
In the Call group box, select From: Block FED_BATCH_CONCENTRATION
and At: 0.00 min.
Click OK to insert the new block.
The new block will be placed inside the Block FED_BATCH_CONCENTRATION
block.
Action
In the Text pane, select the Base Time instruction of the Block
STABLE_DELTA_P block.
In the Instruction box:
Set Breakpoint to 0.00 min.
In the Instructions group box, select Recirc and DeltaP.
In the Parameters field, set Setpoint to an appropriate value, for instance
1.5 bar.
Click Insert.
Action
In the Text pane, select the Base Time instruction of the Block
STABLE_DELTA_P block.
In the Instruction box:
Set Breakpoint to 0.00 min.
In the Instructions group box, select Specials and DeltaP_WatchPar.
UniFlux 10 Instrucciones de funcionamiento 28993587 AE