B Create a method with UNICORN 6 and above
B.2 An example of using the Method Editor
B.2.5 Phase 3: Fed Batch Concentration
Nested block: Trans membrane
pressure (TMP) setup
This block sets the TMP value.
Step
1
2
Instruction: Setting the TMP
value
Step
1
2
3
4
252
Action
In the Text Instructions tab, select the Block: STABLE_DELTA_P instruction
of the Phase: Fed Batch Concentration phase.
In the Instruction Box:
Set Breakpoint to 0.00 min.
In the Instructions: list, select Other and Block.
In the Parameters for field, type an appropriate name, for instance
TMP_START_STOP.
Click Insert.
The new block will be placed inside the Phase: Fed Batch Concentration
phase.
Action
In the Text Instructions tab, select the Base: SameAsMain instruction of the
Block: TMP_START_STOP block.
In the Instruction Box:
Set Breakpoint to 0.00 min.
In the Instructions: list, select Recirc and TMP.
In the Parameters for field, set Setpoint to an appropriate value, for instance
1.00 bar.
Click Insert.