GE Uniflux 10 Instrucciones De Funcionamiento página 235

Tabla de contenido

Publicidad

Step
2
Instructions: TMP alarm
Step
1
2
UniFlux 10 Instrucciones de funcionamiento 28993587 AE
Action
In the Instruction Box, do the following settings:
Set Breakpoint to 0.00 min.
In the Instructions: list, select Alarms and DeltaP_Alarms.
In the Parameters for field, set High Alarm, Low Alarm, High Warn and
Low Warn to appropriate values according to the specifications for the
filter in use.
In the Parameters for field, set Mode to Enabled.
Click the Insert button.
Action
In the Text Instructions tab, select the Base: SameAsMain instruction of the
Phase: Method Settings phase.
In the Instruction Box, do the following settings:
Set Breakpoint to 0.00 min.
In the Instructions: list, select Alarms and TMP_Alarms.
In the Parameters for field, set High Alarm, Low Alarm, High Warn and
Low Warn to appropriate values according to the specifications for the
filter in use or requirements of the intended application (if lower than
filter specifications).
In the Parameters for field, set Mode to Enabled.
Click the Insert button.
B Create a method with UNICORN 6 and above
B.2 An example of using the Method Editor
B.2.3 Phase 1: Method Settings
235

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido