Step
2
3
4
5
6
7
Instructions: Permeate outlets
Step
1
2
3
4
Nested block: Delay to allow the
valves to react
In this block, the 3 second delay block created earlier will be reused to allow all valves
to react.
Step
1
UniFlux 10 Instrucciones de funcionamiento 28993587 AE
Action
In the Instruction Box:
Set Breakpoint to 0.00 min.
•
In the Instructions: list, select Recirc.
•
Set Feed to Open, click Insert.
Set Feed_Drain to Closed, click Insert.
Set IntegrityTestValve to Closed, click Insert.
Set Retentate_Control_Valve_PCV341 to 100%, click Insert.
Set Retentate to Retentate, click Insert.
Action
In the Text Instructions tab, select the Base: SameAsMain instruction of the
Block: FLOWPATH_CONCENTRATION block.
In the Instruction Box:
Set Breakpoint to 0.00 min.
•
In the Instructions: list, select Permeate.
•
Set Permeate_Drain to Closed, click Insert.
Set Permeate to Permeate, click Insert.
Action
In the Text Instructions tab, select the last instruction before the End_Block
instruction of the Block: FLOWPATH_CONCENTRATION block.
B Create a method with UNICORN 6 and above
B.2 An example of using the Method Editor
B.2.5 Phase 3: Fed Batch Concentration
245