B Create a method with UNICORN 6 and above
B.2 An example of using the Method Editor
B.2.5 Phase 3: Fed Batch Concentration
Step
2
3
4
Nested block: Fed Batch
Concentration parameters
Step
1
2
Instruction: Tank level and
hysteresis
Step
1
2
246
Action
In the Instruction Box:
Set Breakpoint to 0.00 min.
•
In the Instructions: list, select Other and Block.
•
In the Parameters for field, select the Block: DELAY_3_SEC block.
Click Insert.
Action
In the Text Instructions tab, select the Block: FLOWPATH_CONCENTRATION
instruction of the Phase: Fed Batch Concentration phase.
In the Instruction Box:
Set Breakpoint to 0.00 min.
•
In the Instructions: list, select Other and Block.
•
In the Parameters for field, type an appropriate name, for instance
•
PARAMETERS_CONCENTRATION.
Click Insert.
•
The new block will be placed inside the Phase: Fed Batch Concentration
phase.
Action
In the Text Instructions tab, select the Base: SameAsMain instruction of the
Block: PARAMETERS_CONCENTRATION block.
In the Instruction Box:
Set Breakpoint to 0.00 min.
•
In the Instructions: list, select Recirc and TankLevel.
•
UniFlux 10 Instrucciones de funcionamiento 28993587 AE