B Create a method with UNICORN 6 and above
B.1 The Method Editor
B.1.2 Method Editor layout
•
•
•
•
Note:
Instructions placed at the same breakpoint will execute simultaneously, while blocks will
execute in the order they occur in the Text Instructions tab.
Changing valve positions, or changing the speed of the pump does not take place mo-
mentarily, but requires a small amount of time. A small time lag, such as 0,10 min, should
therefore be introduced after time-based instructions affecting mechanical components.
224
Breakpoint — Contains a field for setting the breakpoint for the instruction or block,
in the unit currently selected as Base (Time or Volume).
Instructions: — Contains controls for selecting an instruction from a list. The instruc-
tions available are grouped according to function.
Parameters for — Contains parameter fields for the chosen instruction. The param-
eters will vary between the different instructions.
Buttons — There are four buttons: Insert, Change, Replace, and Delete. Change,
Replace and Delete affect the instruction or block currently selected in the Text In-
structions tab. Insert adds a new instruction before, or after (if it has the same
breakpoint as an old instruction), the current selection.
Changing the breakpoint of an instruction using Change will move not
only the instruction, but also all subsequent instructions in the block. Using
Replace moves only the selected instruction.
UniFlux 10 Instrucciones de funcionamiento 28993587 AE