Step
3
4
Instruction: Setting Delta P
watch
Step
1
2
3
4
Set block end time
Change the block end time to 10 minutes to give Delta P some time to stabilize.
Step
1
2
3
UniFlux 10 Instrucciones de funcionamiento 28993587 AE
Action
In the Parameters field:
Set BaseLine to for instance 0.1 bar.
Set MinPeak to 0.00 bar.
Click Insert.
Action
In the Text pane, select the Base Time instruction of the Block
STABLE_DELTA_P block.
In the Instruction box:
Set Breakpoint to 0.00 min.
In the Instructions group box, select Watch and Watch_DeltaP.
In the Parameters field:
Set Test to Stable_Baseline.
Set Minutes to 1.0.
Set Action to END_BLOCK.
Click Insert.
Action
In the Text pane, select the End_Block instruction of the Block
STABLE_DELTA_P block.
In the Instruction box, set Breakpoint to 10.00 min.
Click Change.
A Create a method with UNICORN 5
A.2 An example of using the Method Editor
A.2.5 Block 3: Fed-batch concentration
213