Vercise Gevia™ 16 Contact IPG DFU

Connecting the Stimulator

Dual Lead Connection
Connect Left DBS Extension to port C
1. Fully insert the male end of the DBS Extension into the Stimulator until it stops.
a. Ensure the Stimulator is charged prior to the implantation.
b. Insert the header plug to verify no Setscrews obstruct the socket. Remove the header plug
in order to insert the DBS Extension.
c. Wipe the DBS Extension contacts before inserting.
d. Insert the DBS Extensions into the header. When fully inserted, the tip of the DBS
Extension will slide into the back of the port and the retention sleeve on the DBS Extension
will be located under the Setscrew.
CAUTION: Verify proper DBS Extension insertion by checking impedances before tightening
the Setscrew. Tightening the Setscrew down onto a contact can damage the DBS
Extension.
2. Verify that the retention sleeve on the DBS Extension is located directly under the Setscrew in
the Stimulator header.