EyeSeeCam vHIT Instructions for Use - EN
Page 7
The EyeSeeCam in not intended for operation in oxygen rich environment, flammable anaesthetics,
flammable agents or similar
The goggle should not be worn by patients with strong defective vision and abnormal rare blink. Please
consult a specialist in such circumstances before using the mask on these types of patients.
The system must be switched off before cleaning
When used for longer time in environments temperatures above 30 °C the surface temperature of the
camera unit can exceed 43 °C (max 44 °C). If this feels uncomfortable for the patient turn off the
equipment for a while.
The EyeSeeCam system must be serviced at least once a year. The service must include a safety test.
No parts can be serviced while in use with the patient.
Do not use the equipment if the enclosure is not intact, but have the equipment serviced.
Only personnel with proper training (skilled personnel) should operate the system.
Proper use of this device depends on careful reading of this manual and all additional instructions and
labels
Let the system obtain room temperature before turning it on. Extra time for obtaining room temperature
may be needed to avoid condensation if the system has been moved from a very cold location to a
warmer one.
The patients or test persons should not wear any optical utilities such as glasses or contact lenses.
The patients or test persons should not wear make up around the eyes