EyeSeeCam vHIT Instructions for Use - EN
3.3
USB mode
The EySeeCam vHIT camera must only be connected to a USB 2.0 port. If the computer only has USB
3.0 ports the USB 3.0 controller must be disabled so that the ports will work in USB 2.0 mode.
To downgrade from USB 3.0 to USB 2.0 please go to the BIOS/Config menu and select "USB". Set the
USB 3.0 controller mode to " Disabled" and press F10 to Save and Exit the BIOS Config menu.
3.4
Hardware setup
•
Connect the USB cable to the computer
•
Connect the other end of the USB cable to the EyeSeeCam camera
•
Mount the camera on the goggle using the spherical ball-and-socket joint (see Figure 2.4)
•
Mount the calibration laser on the bridge of the goggle (see Figure 2.4)
•
Connect the computer power supply to a wall outlet, then connect to computer
3.5
Goggle setup
T
he vHIT goggle generally has one camera (monocular) fixed at the top side(s) of the mask. The
camera is held in place mechanically with a spherical ball-and-socket joint. The vHIT goggle has two
spherical sockets: one for the left and one for the right eye. Therefore the camera is interchangeable
between the left and right eyes.
Figure 3-3 EyeSeeCam vHIT goggles
Page 11