Technical specifications

Technical specifications
Operating pressure range
Operating tidal volume range (volume
control modes)
Maximum single fault pressure
Circuit resistance and compliance range
for stated accuracy of monitoring and
control**
Breathing resistance under single
fault***
Maximum flow
Inspiratory trigger (nominal)
characteristics
Expiratory cycle (nominal)
characteristics
Sound pressure level
Sound power level
Alarm volume range
Data storage
Dimensions (L x W x H)
66
Single limb with valve or double limb with valve: 3 to 50 hPa
Single limb with intentional leak: 2 to 50 hPa
CPAP: 3 to 20 hPa
Maximum working pressure limit: 10 to 55 hPa
Forced cycling occurs if the Pressure alarm limit is exceeded.
Adult patient type: 100 to 2500 mL
Pediatric patient type: 50 to 300 mL*
60 hPa (in all modes)
Pediatric patient setting:
Circuit resistance range (circuit with intentional leak): 0 to 8 hPa at 60
L/min
Circuit resistance range (circuit with valve): 0 to 20 hPa at 60 L/min
Circuit compliance range: 0 to 4 mL / hPa
Adult patient setting:
Circuit resistance range (circuit with intentional leak): 0 to 20 hPa at 120
L/min
Circuit resistance range (circuit with valve): 0 to 35 hPa at 120 L/min
Circuit compliance range: 0 to 4 mL / hPa
Pediatric circuit
Inspiration: 2.2 hPa (at 15 L/min), 5.3 hPa (at 30 L/min)
Expiration: 2.4 hPa (at 15 L/min), 5.0 hPa (at 30 L/min)
Adult circuit
Inspiration: 5.7 hPa (at 30 L/min), 8.3 hPa (at 60 L/min)
Expiration: 4.2 hPa (at 30 L/min), 6.2 hPa (at 60 L/min)
220 L/min
Inspiratory trigger occurs when patient flow exceeds trigger setting.
Double limb with valve (flow trigger): 0.5 to 15.0 L/min
Single limb with valve or double limb with valve: 1.6 to 10.0 L/min (in five
steps)****
Single limb with intentional leak: 2.5 to 15.0 L/min (in five steps)
Cycle occurs when inspiratory flow declines to the set percentage of peak
inspiratory flow.
5 to 90%
35 ± 3 dBA as measured according to ISO80601-2-12:2011.
43 ± 3 dBA as measured according to ISO80601-2-12:2011
56 - 85 dBa (in five steps) as measured according to IEC60601-1-8:2012
7 days of high-resolution airway pressure, respiratory flow and delivered
volume (sampled at 25 Hz).
7 days of breath-related therapy data (sampled at 1 Hz).
365 days of statistical data per program.
285 mm x 215 mm x 93 mm
(11.22" x 8.47" x 3.66")