Appendix A: Definitions

Ventilation settings definitions

The available settings will vary with the selection of the ventilation mode. Each mode details the settings
available.
Setting
Apnea Definition
Apnea Interval
(T apnea)
Apnea Response
Circuit Type
CPAP
Cycle
EPAP
Flow shape
Inspiratory duration option
(Insp Duration Option)
Interface type
Interval
IPAP
Magnitude
Manual Breath
Mask Type
Max EPAP
Max PS
Min EPAP
Min PS
P control
Definition
Apnea Definition sets the type of breath which must be delayed for an apnea to be
detected.
Apnea Interval (T apnea) sets the period without breath or spontaneous breath required
for an apnea to be detected.
Apnea Response sets the behaviour of the ventilator when an apnea is detected.
Circuit Type sets whether a Double limb circuit, Single limb circuit with expiratory valve
or Single limb circuit with intentional leak is in use.
Continuous Positive Airway Pressure (CPAP) sets the pressure maintained throughout a
spontaneous breath.
Cycle (also known as Expiratory Trigger) sets the threshold where start of expiration
within a breath is detected.
Expiratory Positive Airway Pressure (EPAP) sets the pressure to be delivered to the
patient during expiration.
Sets the target flow waveform for the delivery of mandatory controlled volume breaths.
Inspiratory duration option (Insp Duration Option) sets whether Inspiration Time (Ti) or
Peak Inspiratory Flow (PIF) is used to configure volume controlled breaths.
Invasive, mask, or mouthpiece
Sigh interval sets the period between sigh breaths.
Inspiratory Positive Airway Pressure (IPAP) sets the pressure to be delivered to the
patient during inspiration.
Magnitude sets the size of the manual or sigh breath delivered relative to the size of the
normal ventilation breath. Separate magnitude settings are available for configuration of
manual or sigh breaths.
Manual Breath sets whether a manual breath is available for delivery.
Mask Type sets the type of mask or in-line vent in use when the circuit type is single
with leak.
Maximum Expiratory Positive Airway Pressure (Max EPAP) sets the maximum pressure to
be delivered to the patient during expiration to maintain upper airway patency.
Maximum Pressure Support (Max PS) sets the maximum pressure support above EPAP
allowed to achieve the Target Va.
Minimum Expiratory Positive Pressure (Min EPAP) sets the minimum pressure allowed to
be delivered to the patient during expiration to maintain upper airway patency.
The Min EPAP should be set to treat any lower airway condition.
Minimum Pressure Support (Min PS) sets the minimum pressure support above EPAP
allowed to achieve the Target Va (iVAPS).
Pressure control (P control) sets the pressure support above PEEP to be delivered during
inspiration for pressure assisted breaths.
Appendix A: Definitions
English
77