8. Install this cable assembly into the appropriate enclosure opening as indicated in Figure 3-16.
3.4.1.2.
PAB Installation
The IND560x-PAB option installs on top of the main board of the IND560x harsh/analog terminal
as shown in Figure 3-28. The following steps describe the process for installing the IND560x-PAB
in the harsh/analog version of the IND560x terminal.
1. Disconnect all existing connections to the IND560x main board. Remove the fastening screws
2. Install the four brass standoffs provided with the IND560x-PAB kit into the locations indicated in
3-22
METTLER TOLEDO IND560x Installation Manual
The blue and white wires will connect to the IND560x-PAB in the procedure described in the
next section.
at the 4 corners and set aside for later use. Remove the main board from the front cover of the
terminal.
Figure 3-27. The standoffs should be secured to the main board with four screws provided in
the IND560x-PAB installation kit.
Figure 3-27: Location of IND560x-PAB Brass Standoffs
64061929
|
11
|
07/2017