English
System Shutdown
1. There are no special procedures for system shutdown. Depress power switch on rear of console to shut down system.
2. To disconnect power cord from mains, grasp ridged portion of connector between thumb and forefinger of one hand.
Gently pull connector away from console. Connector should remove easily. If not, ensure only ridged portion of
connector is pulled. Do not pull on smooth part of connector, as connector will not disengage from console.
Additional Information
End of Life
Return all equipment for proper disposal.
Recommended Manufacturer Service Interval
It is recommended that the equipment be returned to DePuy Synthes Power Tools at a minimum every 9 months for full
product inspection and service.
Console Alerts, Codes, and Displays
Console Indicators
Code
E1, E8 *
E2
E3
E4, E5 *
E6, E7
E9 *
(none)
* If these codes are displayed, return handpiece and console to DePuy Synthes Power Tools for servicing.
** Note: Consoles with software version 3.0 and higher do not generate an "Audible Signal" for Codes E1 – E9. To determine
software version press the irrigation rate buttons at the same time on the console touchpad. The console will display two
digits. Example Only: "30" means version 3.0. On non-irrigation console models, the buttons can be found in the same
location, although they are not marked.
8
Description
System Fault
Handpiece Lock Engaged
Handpiece Stall Warning
Fault in Handpiece
Handpiece Overheat Warning
Handpiece Not Recognized
Handpiece Set to Reverse
Audible Signal **
none
10 slow beeps
10 fast beeps
none
5 second beep
none
3 beeps