Foot Control Operation
Certain optional foot controls have direction and irrigation control switches.
1. See Step 2 of Console Face Panel Operation to set operational direction of handpiece.
2. (Optional) Depress "Foot" button on face panel to activate Foot Control 1, depress again to activate Foot Control 2. Only
one foot control may be active at a time.
Note: Handpiece is fully functional at this time.
3. Press on foot control pedal to start handpiece. Increase pressure on pedal to increase speed of handpiece and release
pressure on pedal to decrease speed of handpiece.
4. (Optional) To change direction of handpiece, depress Directional Control Switch located on top left of foot control for a
minimum of one second. When direction changes, a single beep will sound and LED on console face panel indicates
direction of rotation. Direction can only be changed when handpiece is not running.
5. (Optional) To activate irrigation, depress Irrigation Control Switch on top right of foot control for a minimum of one
second and then step on center of foot control pedal. Upon activation, LED on face panel illuminates; irrigation pump
and handpiece start.
Hand Control Assembly and Operation
Optional hand-control attachment for handpiece allows user to control operation with a lever attached to handpiece.
Warning: Do not expose handpiece to magnets (such as laying handpiece on magnetic drapes) when in hand control
mode, as handpiece operation may occur.
Warning: It is recommended that a craniotome not be used in conjunction with Hand Control. Excessive rotational force on
hand control may move it resulting in handpiece shutting down.
1. Upon console startup Hand Control is OFF. LED is not illuminated.
2. Rotate handpiece until small dimple at proximal end (hose end) of handpiece faces up. Slide hand control over distal
end of handpiece such that notch on hand control slides over dimple at proximal end (hose end) of handpiece. Slide
Hand Control onto handpiece until secure.
3. To activate Hand Control, depress "Hand" button on console face panel. This deactivates Foot Control; Hand Control
LED on face panel illuminates.
4. To prevent inadvertent operation, ensure silver lever on Hand Control is fully inserted in "Lock" position.
5. Extend silver lever from "Lock" position for operation.
6. Apply pressure on Hand Control to operate handpiece.
Note: Handpiece is fully functional at this time.
7. To deactivate Hand Control, depress "Foot" button on console face panel. To remove hand control from handpiece, slide
towards distal end of handpiece and remove.
System Operational Assessment
1. Confirm attachment and dissection tools are properly locked in place.
2. Depress foot pedal to assure you have properly functioning device prior to proceeding.
3. Confirm device functions as expected. If device does not function as expected refer to troubleshooting chart.
Straight Attachments
Angle Attachments
Craniotomes
Perforator Driver
Foot Pedals
EMAX2
EMAX2PLUS
SC2100
SC2101
SC2102
NOTE: DRIVER Duty Cycle 30 Sec ON, 30 Sec OFF for eight cycles.
The recommendations for times of use for the DRIVER attachments have been determined under average load and worst
case ambient air temperature of 85°F (29°C).
Prior to Use Inspection
Visually inspect for any damage to the tube.
Visually inspect for bent or broken drive shaft and for any damage to the tube.
Visually inspect for bent or broken foot.
Visually inspect for overall damage or missing components.
Visually inspect for damage to the electrical cord or connector.
Visually inspect for damage or cracks to the housing or pedal.
Visually inspect for damage to the silicone hose or to the electrical connector.
Connect to console and operate.
The handpiece should operate smoothly.
There is no requirement to operate with attachment or dissection tool.
Visually inspect for damage or cracks to the housing.
Visually inspect for damage to the electrical power cord.
Power the system and inspect for illumination of the LEDs.
If there is an irrigation pump present check for proper function.