Electromagnetic interferences and electromagnetic compatibility
tables
The MIO-CARE equipment has been designed and manufactured according to the
ELECTROMAGNETIC COMPATIBILITY DIRECTIVE 2014/30/EU with the aim of
providing adequate protection from harmful interference when installed in homes
and health establishments.
The equipment does not generate significant radio frequency energy and is
adequately immune to radiated electromagnetic fields. Therefore, it does not
detrimentally interfere with radio-electric communications, electro-medical
equipment for monitoring, diagnosis, therapy and surgery, office electronic
devices such as computers, printers, photocopiers, fax machines, etc. or any
electric or electronic equipment used in these environments, as long as the
equipment complies with the ELECTROMAGNETIC COMPATIBILITY directive.
In general, the use of accessories other than those specified or provided by the
manufacturer could result in increased electromagnetic emissions or decreased
electromagnetic immunity of the MIO-CARE and result in improper functioning.
In any case, in order to avoid any interference problems, it is recommended to use
the therapy equipment enough far away from critical equipment for monitoring
vital patient functions, and to be careful when applying therapy to patients with
pacemakers. In any case it is recommended to use the equipment at least at 3
meters away from televisions, monitors, cellphones or any other electronic
equipment, in particular portable RF communications equipment (including
peripherals such as antenna cables and external antennas) should not be used
closer than 30cm (12 inches) to any part of the device, including the cables
specified by the manufacturer; otherwise, it could lead to degradation of the
performance of the MIO-CARE.
In conclusion, the use of MIO-CARE adjacent to or stacked with other equipment
should be avoided, since it could cause improper functioning. If such use is
necessary, the MIO-CARE and the other equipment should be constantly observed
to verify that they are operating normally.
When the device is used in an environment relatively dry, strong electromagnetic
interferences usually occur. At this time, the device may be affected as follows:
-
the device stops supplying;
-
the device turns off;
-
the device restarts.
The above phenomena do not affect the basic safety and essential performance of
the device, which can be normally used according to the instructions given in this