SHilEy
pHonATE SpEAkinG VAlVE
™
(SSV/SSVo)
description
The Shiley Phonate Speaking Valve is designed to eliminate
the need to use your finger to block the opening of your
trach tube in order to talk. It is a one-way valve that attaches
to the 15 mm connector of your trach tube, including
pediatric sizes. It is available without an oxygen port (SSV)
or with an oxygen port (SSVO).
Contraindications
The Shiley™ Phonate Speaking Valve is not for use with
patients who have laryngeal stenosis, vocal cord paralysis,
severe tracheal stenosis, airway obstruction, respiratory
infection or heavy pulmonary mucus secretions.
How the Shiley
phonate Speaking Valve Works
™
When the Phonate Speaking Valve is attached to the 15 mm
connector or hub of a fenestrated or cuffless tracheostomy
tube, the valve opens when you breathe in, allowing air to
enter your lungs. When you breathe out, the valve closes and
causes air to go up your windpipe and through your vocal
cords, allowing speech.
Speaking Valve
!
WARninG: Ensure that oxygen tubing
does not excessively tug or pull at
speaking valve attachment.
The Shiley Phonate Speaking Valve
should only be used, on the order of a
physician, with patients who are alert
and cooperative.
The Shiley Phonate Speaking Valve
should not be used while sleeping.
The Shiley Phonate Speaking Valve is
not designed to be used with patients
who are ventilator dependent.
Use the Shiley Phonate Speaking
Valve only with a cuffless tube or a
fenestrated cuffed tube. If a cuffed tube
is used, the cuff must be completely
deflated. Do not use with foam cuff
tubes.
CAUTion: Do not attempt to remove the
flexible diaphragm from the valve cap.
Use the Shiley Phonate Speaking Valve
only after secretions have minimized
following tracheostomy.
Pediatric patients should not use the
Phonate Speaking Valve with oxygen
port (SSVO) unless the port cap is not
accessible to the patient.
noTE: The Shiley Phonate Speaking
Valve should only be used on the order
of a physician. Your home healthcare
provider will decide if the Shiley Phonate
Speaking Valve is right for you.
The Shiley Phonate Speaking Valve
should be replaced after 30 days of
normal use.
Replace the Shiley Phonate Speaking
Valve should the flexible diaphragm
become damaged, sticky or malfunction
in any way.
27