9. Troubleshooting and error codes
9.2 Error codes
44
The following describes error codes that may be displayed indicating
an issue with the unit, potential causes, and steps to correct the issue.
Invalid card
No card
Battery low
Motor
overload
Error
Power fault
Potential issues
Program card is not a Symphony card
Program card is incorrectly inserted
Program card is damaged and not
working
Program card is not inserted
Program card is not completely
inserted
Battery needs charging soon, will
beep every 20 seconds (15 minutes of
pumping time remaining from when
the indication started)
Motor is drawing too much current
Potential fault in electronics of gear
control
Potential fault in electronics of power
supply
Error codes
Actions
- Check if the correct program card
is being used
- Check that card is positioned correctly
- Replace program card
- Insert program card
- Push card in all the way
- Be sure power cord is plugged into the
back of the pump
- Contact the local Medela representative if
this happens repeatedly
- Contact the local Medela representative if
this happens repeatedly
- Contact the local Medela representative if
this happens repeatedly