6.
Working with the ELAN 4 electro sys-
tem
6.1
System set-up
WARNING
Risk of injury and material damage due to inadvertent activation of
the applied part!
►
Secure applied parts that are not actually used against inadvertent
activation ("Off" position), see section on locking applied part
("Off" position).
6.1.1
Connecting the accessories
Combinations of accessories that are not mentioned in the present
instructions for use may only be employed if they are specifically intended
for the respective application, and if they do not compromise the perfor-
mance and safety characteristics of the products.
Also note that any equipment connected at the interfaces must demon-
strably meet the respective IEC standards (e.g. IEC 60950 for data process-
ing equipment, IEC/DIN EN 60601-1 for electromedical devices).
All configurations must fulfill the basic standard IEC/DIN EN 60601-1. Any
individual connecting devices with one another is responsible for such
configuration and must ensure compliance with basic standard IEC/DIN
EN 60601-1 or applicable national standards.
►
Follow the instructions for use of individual accessories.
►
In case of questions, contact your B. Braun/Aesculap partner or
Aesculap Technical Service, Address see Chapter 10.
6.1.2
Connecting the power supply
DANGER
Risk of death by electric shock!
►
Connect the product only to a grounded power supply.
Note
The mains voltage must correspond to the voltage indicated on the type
plate at the back of the unit.
►
Plug in the power cord at power cord socket 13.
►
Connect power supply by inserting into a wall socket.
6.1.3
Switching on the control unit
►
Activate the power ON switch 10.
The power ON indicator 9 and indicator light 5 light up.
The control unit 1 always performs a power-on self-test when the unit
is switched on.
If a malfunction is detected, an error message will appear in the
display 2, see system errors.
6.1.4
Switching off the control unit
►
Activate the power OFF switch 8.
The power ON indicator 9, the indicator light 5 and the display with
touch control panel 2 will fade out.
6.1.5
Putting out of operation
Note
The safe and all-pole disconnection of the product from the main power
supply is only guaranteed when the power cord is unplugged.
►
Switching off the product: Activate the power OFF switch 8.
►
Unplug the power cord from the power socket 13.
The operation of the device is safely terminated.
6.1.6
Connecting the ELAN 4 electro to wireless foot control
GA810
The system configuration menu is used to connect the wireless foot con-
trol to the control unit, see Chapter 6.4.4
6.1.7
Connecting the ELAN 4 electro foot control GA808 to the
control unit
Note
The plug connection of the foot control has a yellow coded ring and a filled-
in dot.
►
Position connector of foot control c so that marking b on the connector
is aligned with marking a on the connection socket for foot control 6,
see Fig. 2.
►
Plug the connector for foot control c firmly into the connection socket
for foot control 6.
Fig. 2
Connecting the foot control
Legend
a
Marking for connection socket
b
Marking for connector
c
Connector for foot control
6.1.8
Connecting the ELAN 4 electro motor cable
GA804/GA805/GA806 or ELAN 4 electro drill tool GA844 to
control unit
Note
The motor cable is sterile. Sterile separation occurs at the motor cable.
Note
The plug connection of the motor cable has a blue coded ring and a hollow
dot.
a
b
c
en
9