Fuse Replacement; Technical Service; Accessories/Spare Parts; Elan 4 Electro Motor Cables, Applied Parts And Foot Controls - B.Braun Aesculap GA800 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 72
Malfunction
Oil spray adapter cannot be plugged in Oil spray adapter not compatible
Counterclockwise rotation pusher
cannot be pressed (for GA844).
Speed control safety catch cannot
beactuated (for GA844)
9.2

Fuse replacement

DANGER
Risk of death by electric shock!
Unplug the device before changing the fuses!
Specified fuse set: 2 units IEC 127 – T 6.3 A switching capacity H (1 500 A
at 250 V/50 Hz)
Use a small screwdriver to release the clip on the fuse holder 12.
Pull out the fuse holder 12.
Replace both fuses.
Reinsert the fuse holder 12 in such a way that it audibly snaps into
place.
Note
If the fuses regularly blow, the unit is defective and has to be repaired. see
Chapter 10.

10. Technical Service

DANGER
Danger to life of patients and users in case of malfunctions and/or
failure of protective measures!
Do not perform any servicing or maintenance work under any cir-
cumstances while the product is being used on a patient.
CAUTION
Modifications to medical equipment may result in loss of guaran-
tee/warranty claims and any approvals.
Do not modify the product.
For service and repairs, please contact your national
B. Braun/Aesculap representative.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
ats@aesculap.de
Other service addresses can be obtained from the address indicated above.
Cause
Counterclockwise rotation pusher
jammed/defective
Applied part in secured off posi-
tion
Speed control safety catch
jammed/defective
Detection
Oil spray adapter cannot be
plugged in
Counterclockwise rotation pusher
cannot be actuated.
Safety catch is in "Off" position
Speed control safety catch cannot
be actuated

11. Accessories/Spare parts

11.1 ELAN 4 electro motor cables, applied parts and
foot controls
Art. no.
Designation
GA804
ELAN 4 electro motor cable with hand lever
(from software version 2.00)
GA805
ELAN 4 electro motor cable with hand control
(from software version 2.00)
GA806
ELAN 4 electro motor cable for foot control
GA808
ELAN 4 electro foot control
GA810
ELAN 4 electro wireless foot control (from soft-
ware version 3.00)
GA822
ELAN 4 electro perforator driver
GA824
ELAN 4 electro Lowspeed motor with intra cou-
pling
GA831
ELAN 4 electro sagittal saw
GA832
ELAN 4 electro reciprocating saw
GA833
ELAN 4 electro transverse saw
GA836
ELAN 4 electro micro sagittal saw
GA837
ELAN 4 electro micro reciprocating saw
GA844
ELAN 4 electro drill tool (from software version
3.00)
GA849
ELAN 4 electro craniotome and multifunctional
handpiece (2-ring)
GA860
ELAN 4 electro MIS hand piece
GA861
ELAN 4 electro Standard handpiece (1-ring) L4
GA862
ELAN 4 electro Standard handpiece (1-ring) L7
GA863
ELAN 4 electro Standard handpiece (1-ring) L10
GA864
ELAN 4 electro Standard handpiece (1-ring) L13
GA865
ELAN 4 electro Standard handpiece (2-ring) L7
GA866
ELAN 4 electro Standard handpiece (2-ring) L10
GA867
ELAN 4 electro Standard handpiece (2-ring) L13
GA868
ELAN 4 electro Standard handpiece (2-ring) L17
GA869
ELAN 4 electro Standard handpiece (2-ring) L22
TA014401
Instructions for use of ELAN 4 electro control unit
GA800 (A4 for collection binder)
Remedy
Use the matching oil spray adapter.
Change applied part.
Switch the safety catch to "On" posi-
tion.
Replace the applied part.
en
21

Publicidad

Tabla de contenido
loading

Tabla de contenido