Figure 1: SiphonX® gravitational anti-siphon device (Model SX-200)
W
:
ARNINGS
F
(USA) L
EDERAL
AW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN
R
EAD THE INSTRUCTIONS FOR USE CAREFULLY BEFORE IMPLANTING THE DEVICE
1.
Indications
The SiphonX® gravitational anti-siphon device is designed to control the siphon eff ect
during the treatment of hydrocephalus by shunting Cerebrospinal Fluid (CSF).
W
:
ARNING
The SiphonX® gravitational anti-siphon device must absolutely be used in conjunction
with a CSF shunt valve.
2.
Contra-indications
As it is absolutely essential that the SiphonX® gravitational anti-siphon device is
combined with a CSF shunt valve, the contra-indications for the device are the
following:
-
established or suspected infections along the length of the shunt (meningitis,
ventriculitis, peritonitis, septicemia or bacteremia) or any infection present in
any part whatsoever of the body
-
patients on anticoagulant therapy or presenting with bleeding diathesis
-
ventriculo-atrial shunts in patients with congenital cardiopathies or other
malformations of the cardio-pulmonary system
-
hemorrhagic CSF, as the presence of blood in the system could lead to an
obstruction in the system.
SX-200
However, if the clinical benefi t outweighs these contra-indications, the implantation
.
of a shunt in such cases is performed under the responsibility of an experienced
neurosurgeon. The clinical condition of the patient must, therefore, be subject to
.
increased surveillance.
3.
Description and Operating Principle
of the SiphonX®
Figure 2: The SiphonX® gravitational anti-siphon device
4