Proper Use; Cleaning And Maintenance; Cleaning; Maintenance - Spencer BLUE SPLINT Manual De Uso Y Mantenimiento

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 20
Characteristics
Size (mm)
Weight (grams)
Characteristics
Size (mm)
Weight (grams)
Sizes subject to ± 10 mm tolerances.
9.

PROPER USE

Operators performing immobilisation must be able to select the most suitable type of splint and the most appropriate splint size.
Before and after the application of the splint, it is necessary to check the circulation of the limb to be immobilised.
Applying rigid splints (Blue Splint , Fixo Splint, Urrà)
• Select the most suitable size according to the characteristics of the limb to be immobilised.
• Shape the splint to give it the most suitable shape (only Blue Splint).
• Apply the splint gently to the limb.
• Close the strap bands, ensuring that the tissues of the fractured limb are not disturbed. Urrà must be closed with gauze bandages.
Application of inflatable splints (Air splint)
• Select the most suitable size according to the characteristics of the limb to be immobilised.
• Unzip the splint
• Gently raise the limb and slide the open splint under the limb.
IT
• Zip up the splint and inflate it with the valve nozzle, ensuring that the compression generated by the splint does not harm the tissue. Preferably use the QMX109 pump for
inflation.
• Close the valve opening
• Check pulse and paleness, movement and sensitivity below the area of application as compression may cause reduced tissue circulation.
Application of vacuum splints (Res-Q-Splint, Res-Q-Splint Plus, EMSoft)
EN
• Select the most suitable size according to the characteristics of the limb to be immobilised.
• Lay the splint on the ground and distribute the filling material evenly.
• Gently raise the limb and slide the open splint under the limb.
• Gently wrap the splint around the limb and close the strap bands
• Connect the suction pump to the valve, and begin depressurising while keeping the device adherent to the limb. With EMSoft, use the conical fitting supplied with the splint
DE
to connect the splint to the pump. Once the required rigidity has been achieved, close the valve and adjust the length of the closing straps if necessary.
After use, return the device to atmospheric pressure by opening the valve.
With EMSoft, close the protective cap
Res-Q-Splint and EMSoft splints are compatible with QMX109, QMX120, QMX125,QMX130 pumps.
FR
EMSoft, requires the use of the conical fitting supplied with the splint.

10. CLEANING AND MAINTENANCE

10.1 CLEANING

ES
Failure to carry out the correct cleaning operations could increase the risk of cross-infection due to presence of body fluids and/or residues.
The operator must wear suitable personal protective equipment, such as gloves, goggles, etc. during all checking and cleaning operations.
Clean the exposed parts with water and delicate soap. Never use solvents or stain removers.
Rinse thoroughly with lukewarm water, making sure you have removed all traces of detergent, which may deteriorate or compromise conditions and durability. The use of
high-pressure water should be avoided.
PT
Allow to dry thoroughly before storing. Drying after washing or after use in a wet environment must be natural and not forced. Do not use flames or other direct heat sources.
If blood is visible, oxidise it before washing with water.
Blue splint can be machine washed at 40°C, after removal of the metal core.
If disinfecting, use products that do not have a solvent or corrosive action on materials constituting the device, in addition to being classified as medical-surgical devices. Be
EL
sure to take all precautions to ensure that there is no risk of cross-infection or contamination of patients and operators.

10.2 MAINTENANCE

The device does not require a routine maintenance program, but checks must be made to verify:
• General functionality of the device
HU
• Cleanliness of the device (remember that the failure of cleaning may cause the risk of cross infections)
• Fulfilment of the requirements of the user manual in section Warnings and Specific Warnings.
• Fulfilment of the requirements of the manual in section on Proper use.
No periodic overhaul is foreseen for the device.
RO

10.3 LIFE SPAN

The device, if used as described in the following instructions, has a life span of 5 years from the date of purchase.

11. TROUBLESHOOTING TABLE

SK
PROBLEM
The pump cannot be connected to
the splint
Fabric or belts are torn
10
Res-q-splint
S
M
560x460
490X380X700
750x540x900
295 ± 50
630 ± 50
EMSoft
S
M
750x600
900x700
585
830
CAUSE
Loss of a fitting or improper
use
Normal wear or improper
use.
L
Characteristics
Size (mm)
1050 ± 50
Weight (grams)
L
1100x800
1130
REMEDY
Check that the pump type is suitable for the device in use
Immediately remove the device from service and contact the manufacturer
Res-q-Splint Plus
S
M
735x640
1100x695 ±
1300x720 ±
600 ± 10
850 ± 10
L
1050 ± 10

Publicidad

Tabla de contenido
loading

Tabla de contenido