Metrohm 794 Basic Titrino Instrucciones Para El Uso página 110

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

3.2. Remote control commands
106
Mode.Parameter.Statistics.ResTab.Selected
Mode.Parameter.Statistics.ResTab.DelN
Entries for the statistics calculations.
.Status:
On/off switching. Requirement for statistics calculations is a valid
assignment, see 3.2.2.9.
.MeanN:
Number of individual results for statistics calculations.
.ResTab.Select: Selection of the table for the statistics calculations.
original: Original table. The original table is (again) set up, i.e. any
individual results which have been deleted are reincorporated
in the statistics calculations.
delete n: Single result lines are removed from the statistics calcu-
lation. All results of the corresponding line in the statistics
table are deleted. Specification of the line number in
.ResTab.DelN.
delete all: Clear entire statistics table. The results can not be reac-
tivated.
.ResTab.DelN: Specification of the line number to be deleted.
3.2.2.24. Mode.Parameter.Evaluation.EPC
Mode.Parameter.Evaluation.Recognition.Selected
Mode.Parameter.Evaluation.Recognition.Window.1.LowLim
Mode.Parameter.Evaluation.Recognition.Window.1.UpLim
etc. up to 9 windows
Mode.Parameter.Evaluation.FixEP.1.Value pH:
etc. up to 9 fix EP's
Mode.Parameter.Evaluation.pK
Parameters for DET and MET: Evaluation of the EP's, see page 32.
.EPC: EP criterion in pH, in mV (with U and Ipol) or in µA (with Upol).
.Recognition.Selected: EP recognition.
all: All endpoints found are recognized.
great: Only the largest EP is recognized.
last: Only the last EP is recognized.
window: Only EP's that lie within set windows are recognized.
OFF: The EP evaluation is switched off.
.Recognition.Window.1.LowLim: Lower limit for window in pH, mV (with U
and Ipol) or µA (with Upol).
.Recognition.Window.1.UpLim: Upper limit for window in pH, mV (with U and
Ipol) or µA (with Upol).
Windows are opened until the lower limit is set to OFF. For every
expected EP, an individual window must be set, see page 34.
.FixEP.1.Val: Fix-EP's in pH, mV (for U, Ipol) resp. µA (for Upol). Fix EP's are
evaluated until the setting OFF is found.
original, delete n,
delete all
1...20
DET: 0...5...200
MET pH: 0.1...0.50...9.99
U, Ipol:
1...30...999
Upol:
0.1...2...99.9
all, greatest,
last, window, OFF
pH:
0...±20.00, OFF
U, Ipol:
0...±2000, OFF
Upol:
0...±200.0, OFF
Input range as LowLim
0...±20.00, OFF
U; Ipol: 0...±2000, OFF
Upol: 0...±200.0, OFF
ON, OFF
794 Basic Titrino

Publicidad

Tabla de contenido
loading

Tabla de contenido