Error Messages - Metrohm 794 Basic Titrino Instrucciones Para El Uso

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

3.1. General rules

3.1.4 Error messages

E20
E21
E22
E23
E26
E27
E28
E29
E30
E31
E32
E33
E34
E36
E37
E38
E39
82
Error messages are added to the status messages and separated from them
by the sign ";".
Check exchange unit.
Exit: Mount Exchange Unit (properly) or &m $S.
Check electrode, short circuit.
Exit: Rectify fault or &m $S.
Check electrode, break.
Exit: Rectify fault or &m $S.
Division by zero.
Exit: The error message disappears on next startup or on recalculation.
Manual stop.
Exit: The error message disappears on next startup.
Stop V reached in SET.
Exit: The error message disappears on next startup.
Wrong object call up
Exit: Send correct path for object. Start path at root.
Wrong value or no value allowed.
Exit: Send correct value or call up new object.
Wrong trigger, this trigger is not allowed or carrying-out of action not
possible.
Exit: Send correct trigger (exception: $D) or call up new object.
Command is not possible in active status. Repeat command in inactive
status.
Exit: Send new command.
Command is not possible during titration. Repeat command during the
conditioning phase or in inactive status.
Exit: Send new command.
Value has been corrected automatically.
Exit: Send new command.
Instrument at the end of the titration and sample data is edited; the instrument
at rest or editing during filling.
Exit: &m $S.
RS receive errors:
Parity
Exit: <QUIT> and ensure settings of appropriate parameters at
both devices are the same.
Stop Bit
Exit: <QUIT> and ensure settings of appropriate parameters at
both devices are the same.
Overrun error. At least 1 character could not be read.
Exit: <QUIT>
The internal working-off buffer of the Titrino is full (>82
characters).
Exit: <QUIT>
794 Basic Titrino