794 Basic Titrino
Calibration data can be entered. They are accepted with &Info.CalibrationData
$G. If calibration data are entered, the calibration date is deleted.
3.2.2.69. Info.Checksums
Info.Checksums.MPList
Info.Checksums.ActualMethod
The checksums can be used to identify the content of a file unequivocally,
e.g. files with identical content
have identical results of the checksums. An empty file has checksum "0". The
calculation of the checksums is triggered with $G.
.MPList:
Result of the checksum of the current measuring point list.
.ActualMethod: Result of the checksum of the current method in the working
memory. Identical methods with different method names have the
same results of the checksum.
3.2.2.70. Info.DetermData
Info.DetermData.Write
Info.DetermData.MPList.1.Attribute
Info.DetermData.ExV
Info.DetermData.MPList.1.X
Info.DetermData.MPList.1.Y
for every measuring point
Determination data in hexadecimal format. A measuring point list is available
in mode DET, MET, SET, and MEAS.
Recalculation of the measuring data is triggered with $G.
.Write:
With "ON", the following nodes can be overwritten:
&Info.DetermData.MP.List, &Info.TitrResults.Var.C4X (X = 0...5),
&Info.TitrResults.TempVar.C7X (X = 0...9), and &Mode.Name.
.ExV:
Volume of the exchange unit, with which the determination was
executed
.MPList.1.Attribute: Attribute
.MPList.X: X coordinate, time
.MPList.Y: Y coordinate, volume
3.2.2.71. Info.TitrResults.RS.1.Value
etc., up to .9
Info.TitrResults.EP.1.V
Info.TitrResults.EP.1.Meas
etc., up to .2
Info.TitrResults.Var.C40
etc., up to .C47
.RS: Values of the calculated results.
.EP: Endpoints with DET, MET, SET:
Volume coordinate in mL, e.g. "1.2340"
Measured value coordinate in pH "5.12", mV (with U and Ipol) "-
241" or µA (with Upol) "43.7".
.Var: Various variables. You may overwrite the variables C40...C45, see
3.2.2.95.
3.2 Remote control commands
read only/read+write
read only/read+write
read only/read+write
read only/read+write
read only/read+write
$G
read only
read only
$G
ON, OFF
read only
read only
read only
115