.Req.Buf2:
.Meas.Buf2:
etc.
$G .Assembly.Bur
.Fill:
.ModeDis:
In TIP, its global status as well as the step number (X) is available.
$G .TIP.X
.Inac:
.Req .Id1:
.Id2:
.Id3:
.Smpl:
.Unit:
.Pause:
.Info:
.Mode...:
Status conditions of the global $H:
Status conditions of the global $C:
Status conditions of the global $R:
$R .Mode.XXXX.QuickMeas:
$R .Mode.DET .Inac:
$R .Mode.MET .Inac:
$R .Mode.SET .Inac:
.Cond.Ok:
.Cond.Prog:
$R .Mode.MEAS .Inac:
$R .Mode.CAL .Inac:
$R .Assembly.Bur.ModeDis:
$R .TIP.Inac:
Status conditions of the global $S:
$S .Mode.XXXX.QuickMeas:
794 Basic Titrino
Instrument in the CAL mode, requesting pH of buffer 2.
Instrument in the CAL mode, measures buffer 2.
Buret in filling process
Buret in DIS mode
Instrument at the beginning or at the end of a TIP.
Instrument in the TIP mode, requesting Id1 after start.
Instrument in the TIP mode, requesting Id2 after start.
Instrument in the TIP mode, requesting Id3 after start.
Instrument in the TIP mode, requesting sample size after start.
Instrument in the TIP mode, requesting unit of sample size after start.
Instrument in the TIP mode, in pause.
Instrument in the TIP mode, in info.
Instrument in the TIP mode, working off a submethod. The detailed
status messages of the submethod appear, see above.
The status message of the action which has been held appears.
If the process is held because a monitored limit has been violated, its status
message is $H.Mode.XXX.Titr.
The status conditions of the global $C are identical with the ones of the global
status $G. They appear when the process has been restarted actively from the
status "Hold" ($C, key <meas/hold> or automatically after elimination of an
error).
Quick manual measurement from the initial status in mode XXXX.
Instrument in the DET mode, inactive.
Instrument in the MET mode, inactive.
Instrument in the SET mode, inactive.
Instrument in the SET mode, conditioning, endpoint reached.
Instrument in the SET mode, conditioning, endpoint not reached.
Instrument in the MEAS mode, inactive.
Instrument in the CAL mode, inactive.
Buret in the DIS mode, inactive.
Instrument in TIP, inactive.
Quick manual measurement from the initial status in mode XXXX.
The instrument gives the status from which it has been stopped. The detailed
status information is therefore identical to for the global status $G.
Violation of monitored limits with action "end" give the status message
$S.Mode.XXX.Inac;EYYY.
3.1 General rules
81