3.2. Remote control commands
118
3.2.2.79. Info.ActualInfo.Titrator.CyclNo
Info.ActualInfo.Titrator.V
Info.ActualInfo.Titrator.Meas
Info.ActualInfo.Titrator.dVdt
Info.ActualInfo.Titrator.dMeasdt
Info.ActualInfo.Titrator.dMeasdV
Info.ActualInfo.Titrator.ERC
$Q sends the current values in the following formats:
CyclNo
V(ml)
Meas:
pH
U, Ipol (mV)
Upol (uA)
T (°C)
dVdt (ul/s)
dMeasdt
pH,U,Ipol mV/s
Upol (uA/s)
T (°C/s)
dMeasdV (mV/ul) -
ERC
NV: Not Valid. If in the signal drift is OFF in modes MEAS and CAL, the signal
drift is NV.
OV will be sent for "overrange".
A time frame can be set up from the cycle number and the cycle time (see
3.2.2.82). The cycle number is set to 0 at the start of a method and it is in-
cremented until the end of the method.
3.2.2.80. Info.ActualInfo.MeasPt.Index
Info.ActualInfo.MeasPt.X
Info.ActualInfo.MeasPt.Y
Info.ActualInfo.EP.Index
Info.ActualInfo.EP.X
Info.ActualInfo.EP.Y
$Q sends the last entry into the measuring point list (.MeasPt) or the last entry
into the list of EP's with DET, MET.
.MeasPt.X"165" Volume (DET, MET)
.MeasPt.Y"3.654" Measured value (DET, MET)
.EP.X"1.234" Volume coordinate of the EP
.EP.Y"5.34" Measured value coordinate of the EP
3.2.2.81. Info.ActualInfo.Display.L1
Info.ActualInfo.Display.L2
Lines of the display. The display can be written to from the computer. Pro-
ceed as follows:
Lock the display, see 3.2.2.91.
$Q sends the contents of the corresponding display line.
DET
MET
SET
127
127
127
1.2345
1.2345
1.2345
3.345
3.345
3.6(mV)
-345.6
-345.6
-345.6
-12.5
-12.5
-12.5
-
-
-
-
-
2.5142
0.7957
0.7957
0.7957
0.7957
0.7957
0.7957
-
-
-
-
10.6326 -
34
-
-
read only
read only
read only
read only
read only
read only
read only
MEAS
CAL
127
127
-
-
3.345
3.345
-345.6
-
-12.5
-
25.0
-
-
-
0.7957
0.7957
0.7957
-
0.7957
-
-
-
-
read only
read only
read only
read only
read only
read only
up to 24 ASCII characters
up to 24 ASCII characters
794 Basic Titrino