794 Basic Titrino
UserMeth.List.1.Bytes
UserMeth.List.1.Checksum
for each method
List of the methods in the user method memory with the following
characteristics:
.Name:
Name of the method
.Mode:
Mode
.Quantity: Measured quantity
.Bytes:
Number of bytes of the user memory used by the method
.Checksum: Checksum of the method, see 3.2.2.68.
3.2.2.48. Config.PeriphUnit.CharSet1
Selection of the character set and the graphics control characters of the Ti-
trino.
IBM means the IBM character set following character set table 437 and IBM
graphics control characters. Select 'IBM' for work with the computer.
3.2.2.49. Config.PeriphUnit.Balance
Config.PeriphUnit.Plot
Selection of the balance type and the signal at the analog output.
3.2.2.50. Config.Aux.Language
Selection of the dialog language.
3.2.2.51. Config.Aux.Set
Config.Aux.Set.Date
Config.Aux.Set.Time
Date and time.
Input format of the date: Year-month-day, two-digit, enter leading zeros.
Input format for the time: Hours:minutes, two-digit, enter leading zeros.
Date and time have to be set with &Config.Aux.Set $G just after entry of the
value.
3.2.2.52. Config.Aux.RunNo
Current sample number.
Set to 0 on power on and initialization. After 9999, counting starts again at 0.
3.2.2.53. Config.Aux.AutoStart
Number of automatic, internal starts.
3.2.2.54. Config.Aux.StartDelay
Start delay time in s. During this time, the data of the preceding determination
are retained.
3.2 Remote control commands
Epson, Seiko, Citizen, HP, IBM
Sartorius,Mettler,Mettler AT,
AND,Precisa
U, dU/dt, V, dV/dt, U(rel), T
english, deutsch, francais, espanol,
italiano, portugese, svenska
YYYY-MM-DD
1...9999, OFF
read only
read only
$G
HH:MM
0...9999
0...999 999
111