Consumption current
Current type
Number of phases
Selection of operating voltage
Operation Mode
Protection against electric shock
Protection against harmful water penetration
Refrigeration
Nominal input power
Maximum electric power
Base regulation (ECUT function)
External fuses
Type of fuses
Radiofrequency Output
Radiofrequency output is in conformity with the power indicated on display at a rated load within the
tolerance level of 15% or 5 Watts, whichever is higher.
Modo
Pure (Corte Puro)
Blend 1
Cut
Blend 2
Blend 3
Efeito 1
Efeito 2
Ecut
Efeito 3
Efeito 4
Spray
Coag
Contact
Bipolar
Electrosurgical Unit Waveforms and Basic Frequency
Pure Cut: 480-KHz Sine Wave
Blend 1: Packets of 480-KHz sine wave with a work cycle of 50% and repetition rate of 30 KHz with a damped sine
wave between each packet.
Blend 2: Packets of 480-KHz sine wave with a work cycle of 37.5% and repetition rate of 30 KHz with a damped sine
wave between each packet.
Blend 3: Packets of 480-KHz sine wave with a work cycle of 25% and repetition rate of 30 KHz with a damped sine
wave between each packet.
ECUT Function: Function used together with forms of Cut and Blends. Work method defined by base regulation with
variations of activation periods (TON: 50 ms) and timeout periods (TOFF: 750ms)
Spray Coagulation: 480-KHz damped sine wave with a repetition rate of 60 KHz
Contact Coagulation: 480-KHz damped sine wave with a repetition rate of 60 KHz
Bipolar: 480-KHz sine wave
Adjustment Range
Pure Cut: increase of 1 watt in the whole range (until 200 W)
Blend 1: increase of 1 watt in the whole range (until 150 W)
Blend 2: increase of 1 watt in the whole range (until 150 W)
Blend 3: increase of 1 watt in the whole range (until 100 W)
ECUT: increase of 1 watt in the whole range (until 200 W)
Spray: increase of 1 watt in the whole range (until 100 W)
Contact: increase of 1 watt in the whole range (until 100 W)
Bipolar: increase of 1 watt in the whole range (until 50 W)
Mains power
Minimum power:
Maximum power:
3.9A (127-VACmains) 2.6A (220-VACmais)
AC (alternate)
Two phases
Automatic voltage selection
Non-continuous operation (intermittent – 10-second activation and 30-
second deactivation)
Equipment of CF Type Class I
IPX1 – Protected against vertical water drops
Natural ventilation by convection
38 VA
625 VA
T
: 750 ms and T
OFF
5A / 250V
Glass (model 20AG), 20 mm, fast.
Potência Máx. Declarada
(c/ carga) (Watts)
200
150
150
100
200
200
200
200
100
100
50
Mode
Power (VA) 127V
Stand By
Mode
Power (VA) 127V
Pure Cut
Blend 1
: 50 ms
ON
Carga
(Ohms)
700
700
700
700
700
700
700
700
300
300
100
Power (VA) 220V
20
38
Power (VA) 220V
590
625
451
476
WEM / SS-200E User Manual
Crest Factor
(±20%)
1,9 @ 100W
3,0 @ 75W
3,5 @ 75W
4,3 @ 50W
1,9 @ 100W
3,0 @ 75W
3,5 @ 75W
4,3 @ 50W
6,3 @ 50W
6,3 @ 50W
1,5 @ 25W
Rev. 11