Spray Coagulation
Contact Coagulation
Fuses
In case of fuse burning, replacement should be made with F5A/250V glass fuses.
To replace a fuse, remove the cover of fuse holder located in the back panel by using a screwdriver or
similar tool, and replace the blown fuses with the spare ones stored in the cover of the fuse holder.
Standards applicable to equipment project and development (main standards)
- NBR IEC 60601-1 standard
- NBR IEC 60601-2-2 standard
- NBR IEC 60601-1-2 standard
Standards applicable to equipment manufacture process (main standards)
- NBR ISO 9001 standard
- RDC standard
- EN ISO 13485 standard
Power Curves
300,0
250,0
200,0
150,0
100,0
50,0
Blend 2
Blend 3
Bipolar
Replacement of fuses
– Electromedical equipment – Part 1 – General safety prescriptions
– Electromedical equipment – Part 2 – General safety prescriptions for high-
frequency surgical device
– Electromedical equipment – Part 1 – General safety prescriptions - 2.
Collateral standard: Electromagnetic compatibility – Prescriptions and essays
– Quality management systems – Requirements.
– Good manufacture Practices for Medical Products
– Quality systems – Medical devices – Particular requirements for the application
of ISO 9001
100W
0,0
100
200
300
400
500
600
700
Declared
Pure Cut output power as a function of load
447
474
307
330
350
362
334
405
190
202
200W
800
900
1k
1,1k
1,2k
1,3k
1,4k
Load (Ohms)
1,5k
2k
WEM / SS-200E User Manual