11 – Electromagnetic Compatibility
Manufacturer's guide and statement – electromagnetic emissions
The Microprocessed Electrosurgical Unit model SS-200E is intended for use in the electromagnetic environment
specified below. The user of equipment should ensure that it is used in that environment
Emission essay
RF emissions
ABNT NBR IEC CISPR 11
RF emissions
ABNT NBR IEC CISPR 11
Emissions of Harmonic Currents
IEC 61000-3-2
Emissions due to power/sparkling
fluctuation
IEC 6100-3-3
Manufacturer's guidelines and statement – electromagnetic immunity
The Microprocessed Electrosurgical Unit model SS-200E is intended for use in the electromagnetic environment
specified below. The user should ensure that it is used in that environment
Immunity essay
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transients
/ pulse train ("Burst")
IEC 61000-4-4
Surges
IEC 61000-4-5
Voltage
drops,
short
interruptions
and
voltage
variations
in
electric
power
input
lines
IEC 61000-4-11
Magnetic field at the
power supply frequency
(50/60 Hz)
IEC 61000-4-8
NOTE: U
is the A/C power supply before the application of essay level
T
Conformity
Group 1
Class A
Class A
Conform
Essay level of ABNT
Conformity Level
NBR IEC 60601
± 6 kV by contact
± 6 kV by contact
± 8 kV by air
± 8 kV by air
± 2 kV in power supply
± 2 kV in power supply
lines
lines
± 1 kV in input/output
± 1 kV in input/output
lines
lines
± 1 kV differential mode
± 1 kV differential mode
± 2 kV standard mode
± 2 kV standard mode
< 5% UT
< 5% UT
(> 95% of voltage drop
(> 95% of voltage drop
in UT) per 0.5 cycle.
in UT) per 0.5 cycle.
40% UT
40% UT
(60% of voltage drop in
(60% of voltage drop in
UT) per 5 cycles.
UT) per 5 cycles.
70% UT
70% UT
(30% of voltage drop in
(30% of voltage drop in
UT) per 25 cycles.
UT) per 25 cycles.
< 5% UT
< 5% UT
(> 95% of voltage drop
(> 95% of voltage drop
in UT) per 5 seconds.
in UT) per 5 seconds.
3 A/m
3 A/m
Electromagnetic environment - guide
The Microprocessed Electrosurgical Unit model SS-
200E uses RF power only for its internal functions.
However, its RF emissions are very low and are
unlikely to cause any interference with close electronic
equipment.
The Microprocessed Electrosurgical Unit model SS-
200E is adequate for use in all non-residential facilities
and in those directly connected to the public low-
voltage electric power distributionmains, which power
buildings for domestic use.
Electromagnetic Environment -
Floors should be made of wood,
concrete or ceramics. If floors are
covered by synthetic material,
relative humidity would be at least
30%.
Power supply quality should be
that of a hospital environment or
typical commercial environment.
Power supply quality should be
that of a hospital environment or
typical commercial environment.
Power supply quality should be
that of a hospital environment or
typical commercial environment. If
the equipment user requires a
continued operation during power
shortage, it is recommended that
the equipment is powered by an
uninterrupted power supply or
battery.
Magnetic field at power supply
frequency should be at levels that
are characteristic of a typical place
in a hospital environment or typical
commercial environment.
WEM / SS-200E User Manual
Guidelines
Rev. 11
36