5.9 Terminating ventilation or Demandflow

Important!
Never empty the oxygen cyl-
inder completely. Return the
cylinder for filling while it
still contains residual pres-
sure. This prevents entry of
moist atmospheric air that
can cause corrosion.
60
50
40
CP R
30
20
10
0

5.10 Alarm signals

Stenosis
Disconnection
< 2,7 bar O
2
60
50
40
30
CP R
20
10
0
1. Check the oxygen supply on the pressure reduc-
er gauge. You should change the bottle in good
time, e.g. at below 50 bar, to ensure an ade-
quate operating time.
2. Close the valve of the oxygen cylinder.
3. Switch off MEDUMAT Easy CPR. To prevent the
ventilator being switched off unintentionally, ON/
OFF switch must be kept pressed down for at least
2 seconds until the LEDs in the alarm panel light
up. If audio response is enabled, you should hear
the message: "Close oxygen cylinder".
The alarm panel signals the following alarms:
Stenosis:
Stenosis, or maximum ventilation
pressure P
sive inspirationphases
Disconnection:Disconnection between
MEDUMAT Easy CPR and patient in
two successive inspiration phases
< 2.7 bar:
Drop in oxygen pressure to below
2.7 bar
:
Battery charge inadequate
All the visual alarms are accompanied by an acoustic
alarm.
If the ventilator detects a malfunction during the
self-test after switching on or during continuous
reached in two succes-
max
Using the ventilator
EN
31