Selecting The Display Unit - Heyl Testomat 2000 Instrucciones De Operación

Ocultar thumbs Ver también para Testomat 2000:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Dynamic
analysis start
Filter capacity dependent
control of the analysis interval
▼▲
Start at 40%
Start interval
End interval >
Residual filter capacity (%)
External
analysis start
▼▲
Dynamic analysis start: Depending of the exhaustion of the water
treatment plant, the interval period automatically decreases when the
residual capacity of the plant decreases. This involves the amount of
water used by the plant being determined. The start interval is reset
when limit value 1 or 2 is exceeded or via a signal at input IN1
(regeneration message from the process controller/filter controller).
Selecting dynamic start
In
=>
Press "ENTER" to confirm the selection
The menu on the left appears.
Enter the "
Under "
", enter the start of dynamic operation as a
percentage (%) of the capacity
Under "
(m) that is desired for the start of dynamic operation
Under "
that is desired for the end of dynamic operation
Select the type of reset to the start interval either via "
", "
" or "
Press "ENTER" to confirm the entry
(An asterisk " * " appears at the end of the line)
External analysis start
External analysis start occurs via a contact at the start input.
Note: The current analysis interval can be interrupted by triggering a
contact at the stop input.

Selecting the display unit

It is possible to program the unit of the displayed value. The units dH,
°f, ppm CaCO
and mmol/l can be selected. All the following inputs
3
and displays will then be displayed in the programmed unit.
In
=>
Select the desired unit
Press "ENTER" to confirm the selection
Password protection and basic program
, select
=>
3
" of the plant in m
", enter the analysis interval in minutes
", enter the analysis interval in minutes (m)
"
, select
=>
=>
27