Description of the signal inputs/outputs
NOTE
!
WARNING
▼▲
▼▲
38
Interfaces (optional)
Current interface 0/4-20 mA
Current interface load
The maximum load of 500 Ohms should not be exceeded!
In case of possible faults and when using very long cables (approx.
20 m), a screened cable should be used, if possible.
Installing the interface cards SK910/RS910
Avoid static charges during installation!
Personnel responsible for installation must be appropriately trained
and qualified!
Switch off the instrument.
Insert the plug-in circuit board into the left-hand slot with the
component side on the left. (Contact no.1 is at the top)
Switch on Testomat 2000
Program the desired type of current (0/4-20mA) (for SK910).
Monitoring a measuring point
A printer can be connected to record the analysis results. The
instrument is equipped with a programmable current output for this
purpose (optional 0-20 mA or 4-20 mA).
The example on the left displays the current profile in the 0-20mA
range for one measuring point.
In
=>
Select the desired current range
Press "ENTER" to confirm the selection
In
=>
Select the desired configuration
Press "ENTER" to confirm the selection
®
.
, select
=>
, select
=>